Reads 1,500 patches in 4 minutes
Automatic wavelength calibration
Measures a wide range of charts and substrates
Temperature drift compensation
Automatic QR code recognition and chart alignment
Virtual Fluorescent Standard for non-standard viewing environments
HIGH-SPEED EXECUTION
The MYIRO-9 can read 1,500 patches in just 4 minutes and 2,500 patches in just 7 minutes. With a single scan, users can obtain data for M0, M1, and M2, conforming to ISO13655:2017. This provides improved color accuracy for papers that contain optical brightening agents (OBA), increasing certainty of color reproduction.
BOUNDLESS VERSATILITY
The MYIRO-9’s free-format-function allows it to measure virtually every chart used to calibrate printers and create profiles, providing flexibility, convenience, accuracy, and precision in color calibration. With its ability to measure a wide range of charts and targets, the MYIRO-9 offers unparalleled versatility and compatibility, making it a valuable tool for any printing or color management workflow.
UNMATCHED ACCURACY
The MYIRO-9 is equipped with built-in wavelength calibration and temperature drift compensation to correct for any variations that might affect color measurement accuracy, such as changes in ambient temperature or aging of the instrument’s optical components. Users can expect highly precise and consistent color measurement results, even in demanding color-critical applications. The MYIRO-9 also includes a self-check function, using certified calibration tiles for regular accuracy audits.

MYIRO-9
Illumination/viewing system
45°a: 0°(annular illumination) *1
*1 Conforms to CIE No. 15, ISO 7724/1, DIN5033 Teil 7, ASTM E 1164 and JIS Z. 8722 Condition A for reflectance measurements.
Spectral separation device
Concave grating
Wavelength range
380 ~ 730 nm
Wavelength pitch
10 nm
Half bandwidth
Approx. 10 nm
Measurement area
Approx. Ø 3 mm
Light source
LED
Measurement range
Reflectance: 0 to 150%
Short-term repeatability
Colorimetric: Within σΔ E00 0.05 (Without polarization filter)
* Under Konica Minolta test conditions where a white calibration plate is measured 30 times at 10- second intervals after white calibration
Inter-instrument agreement
Within E00 0.3 (Average of 12 BCRA Series II colour tiles compared to values measured with a master body under Konica Minolta standard conditions; without polarization filter)
Measurement time
Approx. 4 min per 1,500 patches
*Konica Minolta measurement conditions*3
Output item
Spectral reflectance
Measurement conditions *2
M0 (A), M1 (D50) , M2 (A + UV filter), C, ID50, D65, ID65, F2, F6, F7, F8, F9, F10, F11, F12, User-defined light source
Backing condition
White backing, compliant with ISO13655
Interface
USB2.0, 10Base-T/100Base-TX
Power
Specified AC adapter (Input: AC 100 to 240 V, 50/60 Hz; Output: DC 24 V, 5 A)
Size
MYIRO-9 only: W716 x D251 x H159 mm
Auto Sheet Feeder: W601 x D1,158
(when tray is open) x H245 mm
Weight
MYIRO-9 only: Approx. 10.5 kg
Operating temperature/
humidity range
10 to 35°C, 30 to 85% relative humidity with no condensation
Storage temperature/humidity
range
0 to 45°C, 0 to 85% relative humidity with no condensation
Standard accessories
AC adapter, USB cable, Measurement Utility software FD-S2w
Optional accessory
Auto Sheet Feeder FD-A09
Measurable Chart Specifications
Width
45 ~ 330mm
Length
170 ~ 1500 mm
Thickness
0.05 ~ 0.6 mm
Smallest patch size
6×6 mm
Maximum patches per sheet
1,394(A4)
2,928(A3)
*1 Unidirectional illumination for wavelengths of 400 nm and below
*2 M0, M1 and M2 are lighting conditions contained in ISO13655 “4.2.2 Illumination requirements and measurement conditions.”
*3 Chart size: A3, Patch size: 6 x 6 mm, Patch layout: 47 rows x 32 columns